User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HP 6038A:System Power Supply DATE: 26-Sep-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 46 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 ASK- R N P F W 1.003 HEAD * INITIAL CONDITIONS * 1.004 DISP The following test equipment is required: 1.004 DISP 1. Metered vaiac DTB # 2-172 1.004 DISP 2. Transistor Devices DLP 50-150-3000A 1.004 DISP 3. Sensitive Research 20 Amp Shunt DTB # 14-109 1.005 DISP Set the UUT front panel controls as follows: 1.005 DISP LINE ON/OFF ........................................ OFF 1.005 DISP OUTPUT ADJUST ................................ fully CCW 1.006 DISP Connect the System port 1 to the UUT HP-IB connector. 1.007 DISP Remove the UUT rear panel terminal strip cover. 1.008 DISP All connections are made to the rear panel 1.008 DISP terminal strip. 1.009 DISP [32] Connect the UUT and the 34401A as follows: 1.009 DISP [32] 6038A TO 34401A 1.009 DISP [32] OUTPUT (+S) ......................... INPUT-HIGH 1.009 DISP [32] OUTPUT (-S) ........................... INPUT-LO 1.010 DISP Connect UUT AC line cord to Metered Variac. 1.011 DISP Set the variac to 115 VAC. 1.012 DISP Set the UUT front panel controls as follows: 1.012 DISP LINE ON/OFF ......................................... ON 1.013 IEEE [D3000] 1.014 CALL HP 6038A: Sub 1 1.015 CALL HP 6038A: Sub 2 1.016 IEEE CLR; 1.017 HEAD 1.018 DISP Set the UUT front panel controls as follows: 1.018 DISP LINE ON/OFF ........................................ OFF 1.019 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 18 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.